HOW ACCURATE IS THE PULSE OXIMETRY READING IN AN INTENSIVE CARE SETUP?

Authors

  • DR. R. SRIVATSAN Department of Biochemistry, Sri Manakula Vinayagar Medical College and Hospital, Pondicherry, India
  • DR. S. ASMATHULLA Department of Biochemistry, Sri Manakula Vinayagar Medical College and Hospital, Pondicherry, India
  • DR. S. GIRIJA Department of General Medicine, Sri Manakula Vinayagar Medical College and Hospital, Pondicherry, India

Keywords:

Pulse oximeter, SIRS, accuracy, SpO2, SaO2

Abstract

The pulse oximeter remains a valuable tool in the care of intensive care patients, but an awareness of its limitations is important in enhancing the quality of intensive care. We tried to determine the accuracy of pulse oximeter among Systemic Inflammatory Response Syndrome (SIRS) patients in the presence of various factors and also the reliability of predicting SaO2 from subsequent SpO2. Accuracy was analyzed by Bland-Altman analysis. Bias was found to be 2.5%. Subgroup analysis revealed hypoxia and acidosis significantly altered pulse oximeter accuracy. Regression analysis was done to assess reliability of predicting SaOfrom subsequent SpO2. Fluctuations in SpO2 to SaOdifference indicated that SaO2 could not be reliably predicted from SpO2 after a single ABG analysis. A SpO2 above 93% appears necessary to ensure a SaO2 of 90%.

Published

31.12.2015

How to Cite

DR. R. SRIVATSAN, DR. S. ASMATHULLA, & DR. S. GIRIJA. (2015). HOW ACCURATE IS THE PULSE OXIMETRY READING IN AN INTENSIVE CARE SETUP?. International Journal of Pharma and Bio Sciences, 6(4), 1069–1076. Retrieved from https://ijpbs.net/index.php/journal/article/view/4807