METHOD DEVELOPMENT AND VALIDATION OF VALSARTAN AND HYDROCHLOROTHIAZIDE IN PHARMACEUTICAL DOSAGE FORM

Authors

  • V. MASLARSKA Department of Chemistry, Faculty of Pharmacy, Medical University, Sofia, Bulgaria
  • L. PEIKOVA Department of Pharmaceutical Chemistry, Faculty of Pharmacy, Medical University, Sofia, Bulgaria
  • B. TSVETKOVA Department of Pharmaceutical Chemistry, Faculty of Pharmacy, Medical University, Sofia, Bulgaria

Keywords:

Liquid Chromatography, Validation, Valsartan, Hydrochlorothiazide, Tablet dosage form.

Abstract

A simple, rapid and precise method was developed for the simultaneous determination of Valsartan (VAL) and Hydrochlorothiazide (HTZ) in tablet dosage form. Analysis was performed on a C18 (125 x 4 mm, 5 μm) column with acetonitrile: phosphate buffer (pH 3) 45:55 (v/v) as mobile phase, a flow rate 1.0 mL/min, column temperature 30ºC and UV detection at 250 nm. Both the drugs were well resolved on the stationary phase and the retention time for hydrochlorothiazide was 3.124 min and for valsartan 6.469 min. The calibration curves were linear in the concentration range of 5.00-50.00 μg/mL for VAL and 1.0-10.0 μg/mL for HTC. Intra- and inter-day relative standard deviations for both the components were <2.0%. The percentage recoveries obtained for VAL and HCT ranges from 99.28 % to 100.3 %.

Published

31.03.2014

How to Cite

V. MASLARSKA, L. PEIKOVA, & B. TSVETKOVA. (2014). METHOD DEVELOPMENT AND VALIDATION OF VALSARTAN AND HYDROCHLOROTHIAZIDE IN PHARMACEUTICAL DOSAGE FORM. International Journal of Pharma and Bio Sciences, 5(1), 205–211. Retrieved from https://ijpbs.net/index.php/journal/article/view/3032

Issue

Section

Research Articles

Categories