RP-HPLC Method Development And Validation For Simultaneous Estimation Of Sulbactam And Cefoperazone In Dosage Form And In Plasma

Authors

  • K.VINOD KUMAR KMCH College of pharmacy, kalapatti road, Coimbatore-641035
  • J.DHARUMAN KMCH College of pharmacy, kalapatti road, Coimbatore-641035
  • A. KARUNA SREE Raghavendra Institute of Pharmaceutical Education and Research, Anantapur Dt, India-515721

Keywords:

RP-HPLC, Sulbactam, Cefoperazone

Abstract

A simple, precise and accurate RP-HPLC method has been developed for the simultaneous estimation of sulbactam and cefoperazone in dosage form and in plasma using ornidazole as an internal standard. The chromatographic separation was achieved on a phenomenex phenyl hexyl column (250mmX4.6mm,5µm) using a gradient system with mobile phase composed of acetonitrilephosphate buffer(sodium phosphate-20mM)(65:35v/v)pH 3.5 adjusted with orthophosphoric acid at a flow rate of 1 ml/min and detector wavelength190nm.The retention time of sulbactam and cefoperazone was 3.1 and 4.1.The validation of the proposed method was carried out for specificity, linearity, accuracy, precision, robustness and limit of detection and quantitation.The linear dynamic ranges were from 10-50µg/ml for sulbactam and 10-50µg/ml for cefoperazone.Limit of detection and quantitation for sulbactam  were 50ng/ml and 300ng/ml for cefoperazone were 50ng/ml and 400ng/ml respectively. The developed method can be used for routine quality control analysis of titled drugs in combination in formulation and in plasma. 

Published

31.12.2010

How to Cite

K.VINOD KUMAR, J.DHARUMAN, & A. KARUNA SREE. (2010). RP-HPLC Method Development And Validation For Simultaneous Estimation Of Sulbactam And Cefoperazone In Dosage Form And In Plasma. International Journal of Pharma and Bio Sciences, 1(4), 87–92. Retrieved from https://ijpbs.net/index.php/journal/article/view/146

Issue

Section

Research Articles